DUCOM FPT 冷冻点试验器 (+80~-90°C)

Measuring parameters:

Temperature: °C/°F

Measuring Range: +80°C to -90°C (410/2 SA)

Resolution: 0.01°C

Accuracy: ±0.1°C

Repeatability/Reproducibility: as per standard methods or better

Temperature measurement: PT100, Class A

Operating range: Ambient to 32°C; Max. R.H. 80%

Stirrer: A micro-motor drives all the mechanical system 3 coils stirrer made of brass.

Measuring devices:

Freezing Point Principle: According to the methods, the sample is cooled down and stirred. The solid hydrocarbon crystals formation are detected by means of a light beam through fiber optic reflected thanks to a mirror. As soon as crystals are detected, the sample is warmed up until their complete disappearance.

Freezing Point Device: Light pulsed emission on I.R spectrum through a coaxial fibber optic. Coaxial fibber optic equipped with a mirror

Dimensions and weight: 34x60x80 cm (WxDxH), 34 kg 

Electrical supply: 220V+15% / 50 to 60 Hz (or) 115V+15% / 60 Hz