GTI 3991 Aerosol Photometer
Manufacturer: GTI Model: 3991 - Contact
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222)730.39.68
HCM city: (028) 38.119.636
Data Accuracy: 0.0001~100.0% 8.4inch Touch LCD Screen
Dynamic Range: 0.0001µg/L~600µg/L
Accuracy: 1% of Readings(0.01~100%)
Repeatability: 0.5% of Readings (0.01~100%)
Flow Control: 1cfm (28.3L/min)±5%
Zero Building: Automatic or manually
Alarm Modes: Audible, visual and vibratory
Power-Saving Mode: Lower energy consumption
Reporting Modes: Continuous, Summary, Monitoring mode
Print: Built-in thermal printer
ID Encode: ID scanning, data classify and traceable.
Data Output: Storage, USB, Print; Logs can be traced and stored
Data Storage Capacity: 100.000 Groups
Communication: USB is connected with PC continuous scanning testing is available.
GMP Three-Level Privileged Management: Contains three-level permission management, which can set identities and assign usage permissions; it can be done with Log Report
Language: English
Reagent Settings: PAO、DOP、Ondina、PolvethleneGlvcol.Corn Oil、MineralOil、Paraffin Oi, etc.
Sampling Port: Standard internal rectangular sampling port without unnecessary corners
Shell Material: SUS304 stainless steel
Temp. & Humidity: 5~50°C (No condensation, no freezing); 5~95%RH
Storage Temp.: -20C~60°C (Relative humidity below 95%, no condensation)
Power: AC100~240V 50/60Hz or lithium battery (optional with built-in lithium battery pack)
Dimension/Weight: Approx.220x260 (handle 300) x 390mm (W x H x L)/ 11.1kg (photometer) probe and accessories 1kg
Reference complied: NSF49、IEST、ISO14644、EN61010-1:2010、EN61326:2006
What’s included: Photometer, scanning probe, sampling air pipe, manual, certification, power cord, carrying case
Accessories: Lithium battery (built-in), AC/DC, circular sampling port, square sampling port