Hardware:
- Dimensions (W×D×H, mm):
+ Exterior: 1100x570x2460
+ Pass Chamber: 445x370x430
- Material:
+ Exterior, Interior, Pass Chamber: STS 304
+ Front Window: PC (Polycarbonate)
- Weight(kg): 300
Performance:
- Glove Port Diameter (mm): 200
- The Number of Glove Port: 2
- Filter Efficiency: HEPA Filter >99.997% at 0.3 µm
Electricity:
- Power: 220 V, 60 Hz(option: 220 V, 50 Hz)
- Lamp: LED 24 W×1EA
- Power socket: 220 V×2 EA
- Power Consumption 220 V: 210 W
Option: STS 316 Exterior & Interior Material
Lower Stand:
- Support Frame :Height (mm): 900
- BSC Class III manufactured in compliance with international guidelines
Manufactured in compliance with international performance standards from CDC, NIH and AGS to ensure researcher safety.
- Use durable stainless steel 304/316
The durable STS 304/316 is used to round and weld all sides and corners to facilitate cleaning and decontamination of the interior workspace. A valve made of stainless steel allows easy control of air flow and internal pressure.
- High efficiency HEPA Filter
Filter the air with one HEPA Filter and two HEPA Filters, respectively, when entering and exiting the air.
- Convenient filter integrity test
Each filter box is equipped with Upstream, Downstream ports to facilitate filter integrity testing.
- Decontamination Ports
It is possible to select the location of Decontamination ports considering the type of pollutant and the specifications of BSC Class III, and the supply/exhaust pipe connection has been simplified to remove contamination through the coupling port.
- Ability to incorporate a wide range of systems
RTP (Rapid Transfer Ports), Dunk Tank, Autoclave, and Bag-in/Bag-out (BIBO) systems can be easily combined to introduce and remove samples.
Biological Safety Cabinet Class Ⅲ
The Biological Safety Cabinet is a basic safety equipment that is used to protect researchers from pathogenic microorganisms and infectious substances during material handling, tests, and experiments while ensuring a safe laboratory environment
Airflow diagram:
Airflow Characteristics
Due to the nature of the Class III bio-safety workstation where Turbulent Flow is formed, air is introduced through one HEPA Filter and discharged to the outside through two HEPA Filters through the workspace.